Infant incubator (Order Number: IIC0425)
  Two control modes: Air mode and baby mode controlled by micro-computer
  The information can be recorded and checked by the separate LCD display screen
  Humidity and oxygen concentration display and servo control
  >37℃ temperature set function
  The air circle with outlet in both front and back
  Brushless DC motor with low noise
  Double wall hood, operating doors on three sides
  X-ray cassette under the bassinet
  Integrated water reservoir in the shape of drawer which can be autoclaved; Inclination of bassinet is adjustable;
  The height of whole unit can be adjusted
  Failure alarm indication;
  RS-232 connector
Standard configuration
  Main body(including baby compartment, base, controller, I.V.pole and tray), the senor box, skin temperature sensor and vertical height adjustment stand (VHA stand)
Optional Configuration:
  Weighing system, Monitor shelf and Disposable skin temperature sensor
Specification
1. Power supply: AC220V-230V/50Hz or AC110-120V/50-60Hz , 1300VA
2. Mode of control: Air mode and baby mode controlled by micro-computer
3. Air temperature control range: 25℃ -37℃ (override mode 37-39℃ )
4. Baby temperature control range: 34-37℃ (override mode 37-38℃ )
5. Accuracy of skin temperature sensor: ∓ 0.2℃
6. The temperature difference between incubator temperature and average incubator temperature (under steady temperature condition) ≤ 0.5℃
7. Temperature uniformity ≤ 0.8℃ (mattress is on the horizontal position)
8. ≤ 1.0℃ (mattress is on the tilt position)
9. Inclination of the bassinet: ∓ 12 (stepless adjustable)
10. Noise inside hood: 45dB(A)
11. Failure alarm: over-temperature alarm, temperature deviation alarm, sensor alarm, motor
fan failed alarm, power failure alarm, water tank position alarm, system alarm and so on
12. Humidity display range: 0%RH-99%RH
13. Humidity control range: 0%RH-90%RH
14. Accuracy of humidity display : ∓ 5%RH
15. Accuracy of humidity control: ∓ 5%RH
16. Indication range of oxygen concentration: 0%-99%
17. Control range of oxygen concentration: 20%-60%
18. Indication accuracy of oxygen concentration:∓ 2% (oxygen concentration setting value is 25%
and below); ∓ 3% (Oxygen concentration setting value is 25% and above)
19. Control accuracy of oxygen concentration: ∓ 4%
20. Range of weigh: 0-8000g (weight system)
21. Resolution for weight indication: 1g (weight system)
22. Accuracy of weigh indication: ∓ 1% (weight system)
Environmental temperature
  Operating range: 20℃-30℃(Set temperature should be higher 3℃ than the environment temperature)
Ambient air movement rate: 0.3m/s
Package
  Package: each unit is packed as one carton
  Size (Main body):1110mm x 707mm x 883mm
  (Cabinet): 1029mm x 697mm x 826mm
  Gross Weight (Main body): 70Kg
  (Cabinet): 80Kg